Mains powered vaccine refrigerator

VC225ILR

The Dulas VC225ILR is the largest Hot Zone rated large capacity next generation ‘ice-lined’ vaccine refrigerator. A bespoke design for use in the cold chain, it is the result of over 40 years’ experience gained in supplying refrigerators to the most remote corners of the earth.

Bringing Freeze-free™ technology to mains powered refrigerators, the water ‘ice lining’ in the VC225ILR is replaced with our advanced non-corrosive phase change material (PCM), which becomes an ILR with no vaccine freeze risk.

At 184.5 litres vaccine capacity the VC225ILR is ideal for larger health facilities and works with as little as 6 hours of electricity each day and is supplied with a built-in power regulator and voltage stabiliser.

All Dulas refrigerators are supplied with an integrated Beyond ICE3 device fitted at no additional cost. Applying RTM technology is a simple way to improve the reliability of cold chain equipment, enhance maintenance and ensure fewer vaccines are spoiled. Subscription can be activated at purchase or switched on remotely at any point after commissioning the refrigerator.

Features and benefits

Dulas products exceed the Performance, Quality and Safety standards set by the World Health Organisation
With long lasting autonomy you can rely on Dulas products to keep your vaccines safe
Dulas solar refrigerators are battery-free and use an advanced non-corrosive phase change material (PCM) which thaws at exactly +5°C
Stacking baskets for easy stock management and optimum airflow
Dulas products have Grade A User-independent Freeze Protection (UIFP)
All Dulas refrigerators have integrated remote temperature monitoring as standard
climate friendly and CFC-free
for easy handling
Dulas Refrigerators will operate at a stable +2°C to +8°C in an extended ambient operating range of +5°C to +43°C
Renowned for their quality and reliability, the Dulas range of refrigerators are built to perform effectively in challenging environments over long periods of time
to keep vaccines safe

to safeguard equipment from voltage fluctuations.

Technical specification

Model VC225ILR
Performance
Net vaccine storage capacity 184 litre
Gross volume 203 litre
Climate zone Hot zone (+43°C)
Minimum rated ambient temperature +5°C
Holdover time † 94 h 02 min at +43 °C
Hours of electricity required to operate >6 h per 24 h
Refrigerator Dimensions
External dimensions: H x W x D 980 x 1282 x 740 mm
Net weight 147 kg
Shipping Dimensions
Shipping dimensions: H x W x D 1180 x 1355 x 825 mm
Shipping weight 194 kg
Technical Details
Nominal voltage 230 V a.c (120 V a.c version available)
Operating voltage range 82 to 280 V a.c
Nominal frequency 50 Hz
Operating frequency range 45 to 60Hz
Compressor Secop (Danfoss)
Refrigerant R600a
WHO Pre-qualification
PQS reference E003/072
Test procedure E003/RF03-VP.2
Performance specification E003/RF03.2

†Holdover is the amount of time that the refrigerator can maintain the vaccine load within the temperature range of +2 °C to +10 °C without power and at the maximum rated ambient temperature.